8 - Maintaining the instrument 186
Calibration table for nitrogen calibration
Suggestion
The following table is a recommendation for a calibration. The user can customize settings to the
specific analytical task.
Number Sample type
Substance/
Sample weight
Absolute
content
Name Method
2 Conditioning Aspartic acid /
300 mg ±10 mg
RunIn CalibrationLow
1 Conditioning Aspartic acid /
10 mg
RunIn CalibrationLow
2 Conditioning Aspartic acid /
1 mg
RunIn CalibrationLow
20 Calibration
samples
Aspartic acid /
1, 2, 3, 4, 5, 7 ,10,
15, 20, 25, 30, 50,
75, 100, 125, 150,
175, 200, 250,
300 mg
1 mg
0.105 mg N
Aspartic acid CalibrationLow
4 Calibration
samples
Aspartic acid /
350, 450, 550,
650 mg
650 mg
68.38 mg N
Aspartic acid CalibrationHigh