2 Safety
11 Multivapor™ Operation Manual, Version E
Symbol Meaning
Wear protective goggles
Wear protective gloves
Additional user information
Paragraphs starting with NOTE transport helpful information for working with the device / software or
its supplementaries. NOTEs are not related to any kind of hazard or damage (see example below).
NOTE
Useful tips for the easy operation of the instrument / software.
2.5 Product safety
The Multivapor is designed and built in accordance with current state-of-the-art technology, however,
risks to users, property, and the environment can arise when the instrument is used carelessly or
improperly.
The manufacturer has determined residual dangers emanating from the instrument
if the instrument is operated by insufficiently trained personnel.
if the instrument is not operated according to its proper use.
Appropriate warnings in this manual serve to make the user alert to these residual dangers.
2.5.1 Instrument-related hazards
Pay attention to the following safety notices:
!
CAUTION
Risk of minor or moderate burns when handling hot parts.
Do not touch hot parts or surfaces (especially the heating plate with up to 95 °C)
!
WARNING
Death or serious injuries by formation of explosive atmospheres (peroxides) inside the instrument.
Directly withdraw released fumes and gaseous substances by sufficient ventilation at filling
Before operation, check all gas connections for correct installation
Establish inert system atmosphere before processing substances that can form explosive or
reactive gases or powders
Check for proper earth connection to lead off electrostatic charges