Operation ManualPlatformSyncore®093008 en
Syncore Platform102 Safety2.7 Safety proceduresThe regional and local laws and regulations must be obser-ved. It is necessary to wear personal prote
Syncore Platform113 Function3 FunctionSyncore comprises a series of compact, highly versatile instruments suitable for all aspects of parallel synthes
Syncore Platform123.2 Syncore Analyst R-4, R-6 and R-12The Syncore Analyst is a prallel work-up station designed go give high productivity.
Syncore Platform133 Function3.4 Syncore Polyvap R-4 to R-96The Syncore Polyvap comprises a single platform , a rack for samples and fitting glasses
Syncore Platform144 Putting into Operation4.1 Complete and intact deliveryAfter unpacking the delivery, inspect for any signs of damage. It is impor
Syncore Platform154.4 Function test for drive and speed indica-torThere should be no objects of any kind resting upon the base plate or between the b
Syncore Platform164 Putting into Operation Figure 10: Attachment of the condenser unit4.6 Attachment of the condenser unit S and C• Screwtheright
Syncore Platform175 Operation5.1 The single PlatformEnsure that the equipment is properly put into operation according to the recommendations outline
Syncore Platform185 Operation Exceeding the maximum temperatureDo not under any circumstances continue using the unit when the temperature displayed
Syncore Platform195.3 Temperature and speed of rotation1. Press once on the „UP“ or the „DOWN“ button on the tem pe rature control unit. The dis
Syncore Platform205 Operation5.4 Program mode, temperature/time stepsTypical application:• Denitionofthetemperatureandtimeinseveralstages
Syncore Platform21If the UP-key of the time controller is hold until“to 1” appears, the equipment starts the program again using the 1st pro-gram step
Syncore Platform22Figure 15: Display of the set-point temperature for the vacuum connection5.5 Further control parameters5.5.1 Vacuum connection h
Syncore Platform235.6 Pre-set safety settingsStandard pre-set settingsThe Syncore platform allows specific pre-set safety settings to limit the follow
Syncore Platform245 Operation Changing the Pre-Set safety settingsFor the Syncore Analyst we strongly recommend not to raise the maximum temperature
Syncore Platform255.8 Signal-outputsThe Syncore platform is equipped with various signal outlets that significantly extend the functionality of the uni
Syncore Platform265 Operation Figure 19: Display of the post-program running timeThe valve is switched on and off as described below:• The program
Syncore Platform275.11 Signal inputsThe platform also accomodates an analog TTL-signal inlet . An external start or stop signal can be sent to the S
Syncore Platform285 Operation 5.12 Adjusting the eccentricityIntroductionThe unit allows an adjustment of the eccentricity of the horizon tal shakin
Syncore Platform29 5 Operationping position of the base plate also brings about an exact positioning of the access used for manual adjustment of
Syncore Platform3 Table of contentsTable of contentsen, Version E (43 pages) Order codeSyncore Instructions 0930081 Scope of delivery 22 Safety
Syncore Platform30Figure 23: Unbalance compensation5 OperationChecking the settingThe new setting must be checked every time the eccentricity has
Syncore Platform31Adjusting the unbalance compensationNever make any adjustments while the base plate is still moving. Introduction of the tool into t
Syncore Platform325 Operation Checking the settingThe new setting must be checked whenever a change has been made in the unbalance compensation.The t
Syncore Platform335.14 Trouble-shooting: The single platformTrouble-shooting: the single platformObservation Cause CorrectionDisplay does not light N
Syncore Platform346 Maintenance6 Maintenance6.1 CleaningSingle platform• Useaclothdampenedinalcoholtocleanthehousingofthe Syncore single
Syncore Platform356 Maintenance6.3 Routine inspection of the equipment (annually)Temperature control unitThe base plate heater is checked by the use
Syncore Platform366 MaintenanceActivating the calibration mode• TheunitmustbeswitchedOFF.• Toactivatethemodeforcalibrationoftheunit,
Syncore Platform376 Maintenance6.5 Error messages: CalibrationWhenever the progress of the calibration is interrupted by an error, „End“ appears i
Syncore Platform386 Maintenance6.7 Error messages: HardwareError No. Cause Correction1 Hardware ROM BUCHI service agent2 Hardware RAM BUCHI s
Syncore Platform397 Taking out of OperationBefore the unit and its accessories are transported, all hoses must be removed and internal chambers must b
Syncore Platform41 Scope of deliveryFigure 1: BUCHI Syncore Analyst R-6 Voltage
Syncore Platform40Use only genuine BUCHI consumables and spare parts for maintenance and repair to ensure optimum system performance and reliability.
Syncore Platform419 Appendix9 Appendix9.1 Technical dataDimensions of the unit (LxBxH) 480 x 487 x 354 mmWeight 30KgPower connections 230 VAC,
Syncore Platform42English:This equipment has been tested and found to comply with the limits for a Class A digital device, pusuant to both Part 15 of
Syncore Platform439 Appendix9.4 Declaration of conformity
BÜCHI Labortechnik AGCH-9230 Flawil 1 / SwitzerlandT +41 71 394 63 63 F +41 71 394 65 65 www.buchi.com Quality in your hands
Syncore Platform51 Scope of deliveryFigure 2: BUCHI Syncore Polyvap R-241.2 Configuration Syncore Polyvap1 Power system cable PNE, 1.5 m Type CH 10
Syncore Platform61 Scope of deliveryFigure 3: Syncore single platform 1.3 Individual componentsSingle PlatformComponents Order codeSyncore Single
Syncore Platform72 Safety2 SafetyThis unit and all its components have been built in accor-dance with the state of the art and recognized safety rules
Syncore Platform82 Safety2.3 Authorized useThis unit has been designed and built as a laboratory instru ment. The authorized application for the S
Syncore Platform92 SafetySymbol Meaning LocationElemento caliente, superficie caliente General warning • Useofthesingleplatformwithsampleholder