5.6 Pre-set safety settings
Standard pre-set settings
The Syncore platform allows specific pre-set safety settings
to limit the following operating parameters:
• Maximumtemperature,either150°Cor100°C
• Minimumtemperature,either+20°Cor–20°C.
• Rotationalspeedofthebaseplate;either300or600rpm
• Automaticstartofthevortexmovementatthepreviously
set rotational speed or at a fixed rotational speed of 60
rpm.
•InternalbrakeONorOFF
An automatic start of the vortex movement of the base plate
can be of advantage for integration of the unit into fully auto-
mated systems. In such a case, the unit is triggered by an
external signal. A program in the unit can be started with a
starting signal. Whenever the setting „AutoStart of the vortex
movement“ has been selected, the unit uses the speed that
was set last. In this case, no manual control of speed is re-
quired.
The basic settings differ depending on the configuration in
question:
Syncore platform
• Maximumtemperature=150°C
• Minimumtemperature=-20°C
• Maximumrotationalspeed=600rpm
No AutoStart
Syncore Polyvap
• Maximumtemperature=150°C
• Minimumtemperature=+20°C
• Maximumrotationalspeed=600rpm
No AutoStart
Syncore Analyst
• Maximumtemperature=100°C
• Minimumtemperature=+20°C
• Maximumrotationalspeed=300rpm
No AutoStart
5 Operation