BÜCHILabortechnikAG Descriptionoffunction|4
PrepChromC700 25/110
4wayaspirationvalve
The4wayaspirationvalveallowstoperformquaternarygradients.A5
th
port
canbeusedforairpurgeofflashcolumns.
Note
Externalconnectorsallowtopluginasolventlevelsensor(see10.3 Optional
accessories [} page102].Themaximumlengthofallexternalcablesconnected
tothesystemis2meters.
4.3.5 Fractioncollector
Afterthedetectionofthesignal(chromatogram),theeluentattheoutletofthe
column(peaks)canbedirectedtoglasstubesdisposedonracks.Abuiltin
wastevalveallowstochoosebetweenthecollectionortherejection(waste)of
theliquid.
Thecollectorisprovidedwith2stainlesssteelracksfor18mmdiameterglass
tubesasstandard.Thatrepresentsinall128glasstubes.