5|Installation BÜCHILabortechnikAG
42/110 PrepChromC700
NOTICE
Riskofpumpdamage.
Theport"air"onthe4wayaspirationvalveisforairpurgeonly.
u Don’tputtheETFEplugintotheport“air”.
u Installthe4solventsupplytubes(3)intothe4wayaspirationvalve(1)
andtightenthePEEKnuts(2).
HAF700TubingETFE,150cm,2.4mmID,equippedwithspring,4pcs.
NOTE
Watchfortheprotectionspringaroundthesolventsupplytubes.Theymight
havesharpends.
u Installtheinjectionadapter(1)intheportno1ofthe6wayelectrical
valve.
HAF470Injectionadapter