7 Maintenance and repairs
23 F-125 Manual, Version B
7.2 General inspection and cleaning instructions
Check the housing for visible defects (switches, plugs, enclosure etc.) and clean it regularly under
safe conditions with a damp cloth. Wipe off any splashes of aggressive chemicals immediately using
a damp cloth in order to avoid any damage being caused to the coating on the housing. Ethanol as a
cleaning agent is also possible to use.
Cleaning under safe conditions
Switch off the Recirculating Chiller and unplug the power cord.•
Let the system acclimate to ambient temperature completely! ➡
Perform cleaning actions with a damp cloth.•
Regularly clean lamella on front and rear side of the chiller with a damp cloth to remove dust, at •
least once a year.
WARNING
Fire hazard, overheating, or triggering of the over-temperature protection function as a result of a
dirty air intake.
Clean the air intake as needed, but at least 1 x per year.•
Hoses
Check the hoses for wear at least every six months.•
Exchange damaged hoses.•
Cooling media tank
Check prior to use the filling level of the Recirculating Chiller•
Exchange the cooling liquid once a year by means of the drain valve and renew.•
If you use a ethylene glycol / water mixture as cooling media you can drain it to the sink. ➡