5 Putting into operation
33 K-415 Operation Manual, Version B
5.7.2 Hose connections for the TripleScrub
The TripleScrub consists of the neutralization vessel, the adsorption vessel, and the condenser unit.
To establish the hose connections, proceed as follows:
• Put hose connectors GL 14 with seals on all PTFE hoses.
• Connect the emission source to the condenser inlet e.
• Connect the condenser outlet f to the neutralization vessel a.
• Connect the adsorption vessel b to the bypass valve c.
• Cut off a piece of the provided silicone hose and connect the cooling water source to the cooling
water inlet of the condenser g
. Secure it with a hose clamp
.
• Cut off a piece of the provided silicone hose and connect the cooling water outlet of the condenser
h to the waste water system.
1
2
3
4
5
6
7 8
9
a Neutralization stage
b Adsorption vessel
c Bypass valve
d Silencer
e Inlet for fumes from emission source
f Outlet to neutralization vessel (top)
g Cooling water inlet (bottom)
h Cooling water outlet (top)
i Condenser
Fig. 5.12: Hose connections for the TripleScrub
Comentários a estes Manuais