5 Putting into operation
38 K-415 Operation Manual, Version B
5.10 Filling up the prepared washing solution
!
WARNING
Risk of serious chemical burns by corrosives.
Allways use the glass handles of the neutralization vessel to lift or carry it.
Never try to lift or carry the neutralization vessel while holding it on the cover of the neutral-
ization vessel.
To fill up the washing solution, proceed as follows:
Remove the adsorption vessel a by gently pulling it upwards. (If available store it on the tray f.)
Loosen the hose connector GL 14 b at the neutralization vessel and pull off the hose. (If available
store it on the tray f.)
Take the neutralization vessel with the attached cover away from the instrument. Lift or carry the
neutralization vessel solely by using the glass handles h of the vessel.
Remove the cover of the neutralization vessel c togehter with the attached swirl discs d.
Carefully pour the prepared washing solution into the neutralization vessel and fill up to the blue
wavy optimal filling mark g.
Reassemble the instrument in the reverse order.
NOTE
Make sure the cover is pressed firmly against the neutralization vessel to ensure tightness of the
system!
5
1
2
3
4
6
7
8
a Adsorption vessel
b Nut GL 14
c Cover with swirl discs
d Washing insert
e ON/OFF switch
f tray
g filling mark
h glass handles
Fig. 5.15: Filling up the washing solution
NOTE
For information on how and when to exchange the washing solution, see chapter 7.6.