BÜCHILabortechnikAG Operation|6
OperationManualInterfaceI100 23/42
6.2 Performingdistillation
u SwitchontheI100atthemasterswitchontheright.
Dependingontherequirements,theInterfaceI100maybeoperatedinanyofthefol
lowingmodes:
Continuousmode(seeChapter6.2.2"Continuousmode",page24)
Manualmode(seeChapter6.2.1"Manualmode",page23)
Timermode(seeChapter6.2.3"Timermode",page25)
6.2.1 Manualmode
u PresstheSetbutton.
1
1
1
2
Fig.13:Settingtherequiredpressure
u Usingthearrowbuttons,setthespecifiedpressure(1)andthehysteresis(2)and
pressOKtoconfirmineachcase.
NOTE
Thehysteresisisthetolerancebywhichthepressuremayfluctuatebeforeitiscor
rectedbythepump.
u PresstheSTARTbutton.
Thevacuumpumpswitchesonandrunsuntilthesetpressureisreached.There
afterthepumpwillintermittentlystartupautomaticallytomaintainthepressure.