6 Operation
110
K-375/376/377 Operation Manual, Version B
Result units
Create a new unit with NEW and store it under a
new, unique name.
Select a base unit for the new unit calculation
Selectable are: g N/L, % N, g Pr/L or % Pr
(N = nitrogen content, Pr = protein content)
Enter a correction factor and/or correction offset
if necessary and select the preferred number of
decimal places.
(The results are calculated following the linear equa-
tion y= a·x+b where a is given by the correction
factor and b by the correction offset.)
Dosage volume in status view
Set the increment in [mL/click] for the direct
dosing of H2O, NaOH and H3BO3 within the status
view. (Per click on the corresponding dose button,
the specified amount in mL will be dosed.)
Vista de página 109