7 Maintenance
141 K-375/376/377 Operation Manual, Version B
1
To adjust the distance between the sample
tube holder and the splash protector seals
for an optimum contact pressure,
slightly loosen the two screws
insert the gauge into the gap
between the holder and the housing
(insert the gauge according to the
condition of the seal – for a new seal
insert the thick end of the gauge, for
a worn sealing use the thin end, or
the middle part)
move the sample tube holder
upwards against the gauge and
tighten the two screws on the holder
remove the gauge
insert a sample tube and test the
connection of the sample tube for
tightness
if necessary repeat the adjustment
using another step of the gauge
Fig. 7.4 Adjustment of the sample tube holder
NOTE
If the adjustment of the sample tube holder does not remedy the problem, it may be necessary to
exchange the sample tube holder and/or the splash protector seal.