6 Operation
75
K-375/376/377 Operation Manual, Version B
Pump calibration
Select the pump to be calibrated (H
2O, NaOH or
H3BO3).
Enter the target “Dose volume”, e.g. 50 mL.
Press START to start the calibration procedure.
Measure the actually dosed volume and enter it as
calibration volume in the displayed screen. Repeat
the calibration procedure, until the measured and
the dosed volume correspond.
An acceptable difference at 50 mL is ±5 mL.
NOTE
H
2O and NaOH can be dosed into the sample
tube and then poured into a graduated cylinder for
measurement.
The H3BO3 can be dosed directly into the receiving
vessel and then poured into a graduated cylinder.
Sampler functions
The actions “Move to zero position” and “Move to
service position” are available for both sampler types
(1-tray and 2-tray). For a 2-tray sampler the arm is
moved to the corresponding zero or service position
of the tray, the arm is actually positioned in.
For a 2-tray sampler it is also possible to move the
arm of the sampler with “Move to tray A” from tray B
to the zero position of tray A and vice versa.
Press START to move the arm to the selected posi-
tion.
Vista de página 74
1 2 ... 70 71 72 73 74 75 76 77 78 79 80 ... 159 160

Comentários a estes Manuais

Sem comentários

APC Síťové přepínače uživatelské příručky

Začínající na S

Klepnutím na abecedu níže přejděte na úplný seznam modelů začínajících tímto písmenem

Modely Typ Dokumentu
SU042-1 3000 VA
Uživatelský manuál   Manual do Usuário do Chaveador Redundante, 15 stránky
Uživatelský manuál   APC SU042-1 3000 VA User`s manual, 23 stránky
Uživatelský manuál   APC SU042-1 3000 VA User`s manual, 17 stránky
SYMMETRA 990
Uživatelská příručka   APC SYMMETRA 990 User`s guide, 36 stránky
Uživatelský manuál   Symmetra MW II 400-600 W Internal Bypass, 36 stránky
Specifikace   APC SYMMETRA 990 Specifications, 4 stránky