8 Troubleshooting
147
K-375/376/377 Operation Manual, Version B
10‘302 Cooling water flow too low Assure higher flow rate or check parameters in
Settings/Peripherals/Cooling water settings
10‘303 Low pressure during distilling System pressure is below 150 mbar. Check for leaks
or call service.
10‘311 Pump H2O has no current Faulty water pump. Exchange pump or call service.
10’312 Pump NaOH has no current Faulty NaOH pump. Exchange pump or call service.
10’314 Pump H3BO3 has no current Faulty boric acid pump. Exchange pump or call
service.
12’001 Valve steam (Y1) out of order Faulty valve or wire harness. Call service.
12’002 Valve cooling water in (Y5) out of order Faulty valve or wire harness. Call service.
12’003 Valve sampler steam (Y6) out of order Faulty valve or wire harness. Call service.
12’004 Valve sampler transfer (Y7) out of order Faulty valve or wire harness. Call service.
12’005 Valve 5 (not used) out of order Faulty valve or wire harness. Call service.
12’006 Valve waste sample tube (Y2) out of order Faulty valve or wire harness. Call service.
12’007 Valve aspiration in (Y3) out of order Faulty valve or wire harness. Call service.
12’008 Valve receiver (Y4) out of order Faulty valve or wire harness. Call service.
12’009 Valve H2O injection (Y8) out of order Faulty valve or wire harness. Call service.
12’010 Valve H2O sample tube (Y9) out of order Faulty valve or wire harness. Call service.
12’011 Valve waste receiver (Y10) out of order Faulty valve or wire harness. Call service.
13’001 27V power supply overcurrent Faulty electronic board. Call service.
13’002 Ventilator power overcurrent Fan short circuit. Call service.
13’003 Ventilator electronics blocked Check for blockage or call service
13’004 Ventilator interior blocked Check for blockage or call service
14’001 Titrator not ready Check if all cables are plugged to the titrator, restart
the system or call service.
14’002 Titrator information (version) Call service
14‘003 Titrator not started Titrator error. Check function in System Preparation/
Buret function. Call service.
14‘004 Titrator not started, pH value too low pH value is below the set endpoint. Verify the elec-
trode, dosing unit and boric acid.
14‘005 Titrator not started, pH value too high pH value is above the set endpoint. Verify the elec-
trode, dosing unit and receiving solution.
14’006 Wrong titration direction Ensure pH electrode is imersed in receiving solution
and verify if the right titration solution is used.
14’007 Titration speed over specification Overtitration. Use lower concentrated titration solution
or reduce titration speed.
14’008 Titration speed over specification Overtitration during back titration. Use lower concen-
trated titration solution or reduce titration speed.
14’010 Titrator module could not create service 11 Distillation unit needs to be switched off and on
14’011 Titrator module could not create service 21 Distillation unit needs to be switched off and on
14’012 Titrator module could not create service 41 Distillation unit needs to be switched off and on
14’013 Titrator module could not create service 3 Distillation unit needs to be switched off and on