8 Troubleshooting
82 R-220 SE Operation Manual, Version F
Error Addition error indicator on display Possible cause Remedy
E12 ‘FLOW’ message appears. Cooling medium flow insufficient. Increase cooling water flow and
press the “Aerate” button to confirm
the error message.
Adjust sensor.
Check cable and plug of the flow
sensor.
Exchange sensor if defective or run
evaporation without flow sensor.
Call service if problem is persistent.
E13 ‘LEVEL 1’ message appears. Filling level reached. Empty receiving vessel 1 and press
the “Aerate” button to confirm the
error message.
Check cable and plug of the flow
sensor.
Exchange sensor if defective or run
evaporation with user controlled
filling level.
Call service if problem is persistent.
E14 ‘LEVEL 2’ message appears. Filling level reached. Empty receiving vessel 2 and press
the “Aerate” button to confirm the
error message.
Check cable and plug of the flow
sensor.
Exchange sensor if defective or run
evaporation with user controlled
filling level.
Call service if problem is persistent.
E26 Heating bath icon flashes. Bath temp. sensor shortcut. Restart the unit.
Call service if problem is persistent.
E27 Condenser icon flashes. Vapor temp. sensor shortcut. Switch off the unit. Check sensor
cable and plug.
Exchange sensor if defective or run
evaporation without vapor temp.
display.
Call service if problem is persistent.
E28 Condenser icon flashes. Cooling water temp. sensor
shortcut.
Switch off the unit. Check sensor
cable and plug.
Exchange sensor if defective or run
evaporation without cooling medium
temp. display.
Call service if problem is persistent.