5 Putting into operation
30 V-700/710 Operation Manual, Version E
5.4.2 Silencer
Install the silencer to the pump outlet or to the secondary
condenser outlet by screwing it directly onto the GL14 thread.
NOTE
Make sure to mount the silencer in such a way, that the marking
on the silencer has the orientation as shown on the picture on
the left. The orientation is vital to guarantee a correct functio-
ning of the silencer.
Normally, the direction marking on the silencer is not colored.
Fig. 5.4: Silencer
5.5 Woulff bottle or valve unit
To install the Woulff bottle, proceed as follows:
• Screw the corresponding holder to the pump.
NOTE
You can use the allen key integrated in the top cover of the
pump, see also Fig. 4.2.
Comentários a estes Manuais