093091 en
3 Technical data10 V-700/710 Operation Manual, Version ETable 3-1: Accessories Vacuum Pump V-700/710Product Order numberSet hose connections (not on
3 Technical data11 V-700/710 Operation Manual, Version EOrder number:0 7x3x xVacuum Controller V-855 100 V–240 V 50/60 HzOrder number:0 7x4x xEasyVac
3 Technical data12 V-700/710 Operation Manual, Version E3.1.3 Two controllersOrder number:0 7x6 2xTwo vacuum controllers V-850, two valve unitOrder
3 Technical data13 V-700/710 Operation Manual, Version E3.1.5 Condensation assemblyOrder number:0 7x x x1Secondary condenser and insulationOrder num
3 Technical data14 V-700/710 Operation Manual, Version E3.2 Technical data overviewTable 3-2: Technical dataVacuum Pump V-700 Vacuum Pump V-710Dimen
3 Technical data15 V-700/710 Operation Manual, Version E3.4 Solvent tableTable 3-4: Solvent tableSolvent Formula Molar mass in g / molEvaporation en
4 Description of function16 V-700/710 Operation Manual, Version E4 Description of functionThis chapter explains the basic principle of the instrumen
4 Description of function17 V-700/710 Operation Manual, Version E4.1.3 Rear view4456327891 1010a Power supplyb Fan for pump coolingc Pump outletd
4 Description of function18 V-700/710 Operation Manual, Version E4456327891 1010a Power supplyb Fan for pump coolingc Pump outletd Knu
4 Description of function19 V-700/710 Operation Manual, Version E4.1.4 Gas ballastThe gas ballast serves to aerate the system (membrane and valves)
4 Description of function20 V-700/710 Operation Manual, Version E4.2.2 Rear connections of the vacuum controller123456712811 910a Connection for
4 Description of function21 V-700/710 Operation Manual, Version E4.3 Vacuum Module V-801 EasyVacThe Vacuum Pump V-700/710 can be used for the automa
4 Description of function22 V-700/710 Operation Manual, Version E4.3.1 Front view EasyVac123456Fig. 4.8: Front view EasyVaca Detection start of dist
4 Description of function23 V-700/710 Operation Manual, Version EThe LEDs indicate the operating state of the system:c LED ON/OFF• Permanent light:
4 Description of function24 V-700/710 Operation Manual, Version E4.4 Vacuum Module V-802 Lab VacThe Vacuum Pump V-700/710 can be used together with
4 Description of function25 V-700/710 Operation Manual, Version EThe LEDs indicate the operating state of the system:c LED ON/OFF• Permanent light:
4 Description of function26 V-700/710 Operation Manual, Version E4.7 Connections to the secondary condenser123123a Outlet for exhaust gasesb Outlet
5 Putting into operation27 V-700/710 Operation Manual, Version E5 Putting into operationThis chapter describes how the instrument is installed and g
5 Putting into operation28 V-700/710 Operation Manual, Version E5.3 Electrical connectionsATTENTIONMake sure that the voltage on the socket correspo
5 Putting into operation29 V-700/710 Operation Manual, Version E5.4 Silencer or secondary condenser5.4.1 Secondary condenserTo install the secondar
Table of contents3 V-700/710 Operation Manual, Version ETable of contents1 About this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5 Putting into operation30 V-700/710 Operation Manual, Version E5.4.2 SilencerInstall the silencer to the pump outlet or to the secondary condenser o
5 Putting into operation31 V-700/710 Operation Manual, Version E• Introduce the Woulff bottle to the holder.• Screw the GL-14 nut of the Woulff b
5 Putting into operation32 V-700/710 Operation Manual, Version E5.7 Complete system connection with one Rotavapor and one vacuum controller123456789
5 Putting into operation33 V-700/710 Operation Manual, Version E5.8 System pump with two controllers on two RotavaporFig. 5.8: Complete system conne
5 Putting into operation34 V-700/710 Operation Manual, Version E5.9 System pump with two controllers on one pumpOption 1Option 1 is a variant with b
5 Putting into operation35 V-700/710 Operation Manual, Version EThe following installation instruction is valid for both variants, although the contr
5 Putting into operation36 V-700/710 Operation Manual, Version E Commissioning option 1 (parallel)To install the controllers proceed as follows:• 1.
5 Putting into operation37 V-700/710 Operation Manual, Version ECommissioning option 2 (opposite)To install the controllers proceed as follows:• 1.
5 Putting into operation38 V-700/710 Operation Manual, Version ECablingHow to connect the cables:1. Use the two green power supply cables to connect
6 Operation39 V-700/710 Operation Manual, Version E6 OperationThis chapter explains the operating elements and possible operating modes. It gives inst
Table of contents4 V-700/710 Operation Manual, Version E4.6 Vacuum connections to the valve unit or Woulff bottle. . . . . . . . . . . . . . . . . 25
6 Operation40 V-700/710 Operation Manual, Version E6.2 Selecting the distillation conditionsTo achieve optimal distillation conditions, the distillat
6 Operation41 V-700/710 Operation Manual, Version E6.3 Optimizing the distillation conditionsDepending on the solvent being distilled the distillatio
7 Maintenance and repairs42 V-700/710 Operation Manual, Version E7 Maintenance and repairsThis chapter gives instructions on all maintenance work to
7 Maintenance and repairs43 V-700/710 Operation Manual, Version E7.4 Valve head7.4.1 CleaningIf the end vacuum of 10 mbar is not reached anymore and
7 Maintenance and repairs44 V-700/710 Operation Manual, Version E• Remove the two rubber parts at the valve in- and outlet.• Loosen the black tub
7 Maintenance and repairs45 V-700/710 Operation Manual, Version E11• In case the valve has to be replaced, loosen the two screws a by means of the
7 Maintenance and repairs46 V-700/710 Operation Manual, Version E• When all replacement work is done, reas-semble the pump head in the reverse orde
8 Troubleshooting47 V-700/710 Operation Manual, Version E8 TroubleshootingThis chapter helps to resume operation after a minor problem has occurred wi
9 Shutdown, storage, transport and disposal48 V-700/710 Operation Manual, Version E9 Shutdown, storage, transport and disposalThis chapter instructs
9 Shutdown, storage, transport and disposal49 V-700/710 Operation Manual, Version E9.3 Health and safety clearance formDeclaration concerning safety
1 About this manual5 V-700/710 Operation Manual, Version E1 About this manualThis manual describes the vacuum pump V-700 and V-710 and provides all
10 Spare parts50 V-700/710 Operation Manual, Version E10 Spare parts and accessoriesThis chapter lists spare parts, accessories, and options includi
10 Spare parts51 V-700/710 Operation Manual, Version ETable 10-1: Spare parts (cont.)Product Order numberControl cable between vacuum controller and
10 Spare parts52 V-700/710 Operation Manual, Version ETable 10-1: Spare parts (cont.)Product Order numberWoulff bottle complete (with R-210/215 and V
10 Spare parts53 V-700/710 Operation Manual, Version ETable 10-1: Spare parts (cont.)Product Order numberSet of 4 valves, including O-rings and screw
11 Declarations and requirements54 V-700/710 Operation Manual, Version E11 Declarations and requirements11.1 FCC requirements (for USA and Canada)E
11 Declarations and requirements55 V-700/710 Operation Manual, Version E11.2 Declaration of conformity
BÜCHI Labortechnik AGCH-9230 Flawil 1 / SwitzerlandT +41 71 394 63 63 F +41 71 394 65 65 www.buchi.com Quality in your hands
2 Safety6 V-700/710 Operation Manual, Version E2 SafetyThis chapter points out the safety concept of the instrument and contains general rules of beha
2 Safety7 V-700/710 Operation Manual, Version E2.4 Warning notices used in this manualWARNINGGenerally, the triangular warning symbol indicates the p
2 Safety8 V-700/710 Operation Manual, Version E2.6 General safety rulesResponsibility of the operatorThe head of laboratory is responsible for traini
3 Technical data9 V-700/710 Operation Manual, Version E3 Technical dataThis chapter introduces the reader to the instrument specifications. It contai